Highly compact and cost-effective. Micro-sampling from whole blood (CBC : 30 µL - DIFF: 53 µL). Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS Technology.
Features
- Throughput: Up to 60 samples/hour
- Reagents: Only 4 onboard reagents and 1 diluent
- Perfect differentiation of the 5 WBC sub-populations with DHSS Technology
- 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.
- Basophils counted through specific channel
- High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY) and Large Immature Cells (LIC)
Specifications
Technical
Weight |
77 lbs / 35 kg |
Power supply |
100 to 240 V (± 10%) - 50 to 60 Hz |
Dimensions |
20.3 x 17.5 x 19 in. / 51,5 cm x 44,5 cm x 48 cm |
Printer |
Laser |
Throughput |
60 tests/hour |
Data management |
68 331 microprocessor |
Memory |
Last result |
Screen |
16 lines LCD screen |
Samples
60 samples/hour (open tube) |
|
Parameters (20) |
WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV |
Graphs |
High-definition LMNE matrix |
Blood Volume |
CBC mode: 30 µL |
Stability |
48 hours post-draw stability |
Operating sample mode |
Open tube |
Identification |
Alphanumeric. Bar-code reader (optional) |
Results
Parameters |
Precision (%CV) |
Linearity limits |
WBC |
< 2.0 |
0-120 x 10³/µL |
RBC |
< 2.0 |
0-8 x 106/µL |
HGB |
< 1.0 |
0-24 g/dl |
HCT |
< 2.0 |
0-67% |
PLT (whole blood) |
< 5.0 |
0-1900 x 10³/µL |
PLT (PLT concentrate) |
< 5.0 |
0-2800 x 10³/µL |